The Baxter 6301 is a dual-channel volumetric infusion pump.
Programming Modes: ml/hr, volume/time, macro and micro rate
The incremental flow rate can deliver 1-99.9ml/hr. in increments of 0.1ml per hour and from 1-1999ml/hr in increments of 1ml
VTBI Range: 1.0-99.9 mL in 0.1 mL increments or 1-9999 mL in 1 mL increments for both primary and secondary.
Ramp PDP Parameter Ranges: Total VTBI: 100-9,999 in 1 mL increments.
Delay Time: 00:00 - 08:00 hours in minute increments.
Run Time: 01:00 - 16:00 hours in minute increments.
Ramp Up: 00:00 to 45% of the Run Time, in minute increments.
Ramp Down: 00-00 to the Run Time minus Ramp Up, in minute Increments.
Air-in-Line Detection: Factory set to NORM, which causes the pump to alarm on air bubbles approximately 75 uL or larger. The MIN setting causes the pump to alarm on air bubbles approximately 50 ul or larger.
Dimensions:TYPE: Infusion Pump
MANUFACTURE: Baxter
MODEL: 6301
DIMENSIONS: 13” W X 5.1”D X 11.4”H
WEIGHT: 17.9 LBS